Magnetrol 2xx Dokumentacja

Przeglądaj online lub pobierz Dokumentacja dla Pomiary, testowanie i kontrola Magnetrol 2xx. Magnetrol 2xx Specifications Instrukcja obsługi

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj

Podsumowanie treści

Strona 1 - Level Transmitter

Installation and Operating ManualMagnetostrictiveLevel TransmitterSIL 2Enhanced Model 2xxSoftware v3.x

Strona 2 - WARNINGS

1046-648 Jupiter Magnetostrictive TransmittersThe default password installed in the transmitter at thefactory is 0 (password disabled). The last step

Strona 3 - Magnetostrictive Transmitter

46-648 Jupiter Magnetostrictive Transmitters111.6.4.1 Measurement Type: Level OnlyDisplay Action Comment1StatusLevel% OutputLoopTransmitter Display Tr

Strona 4 - 1.0 Installation

1246-648 Jupiter Magnetostrictive Transmitters1.6.4.1 Measurement Type: Level Only (cont.)Display Action Comment24DispFact(select)Select Yes to displa

Strona 5 - 1.3.2 Equipment and Tools

46-648 Jupiter Magnetostrictive Transmitters131.6.4.2 Measurement Type: Interface OnlyDisplay Action Comment1StatusIfcLvl% OutputLoopTransmitter Displ

Strona 6 - 1.4.1 External

1446-648 Jupiter Magnetostrictive Transmitters1.6.4.2 Measurement Type: Interface Only (cont.)Display Action Comment24DispFact(select)Select Yes to di

Strona 7 - Figure 2

46-648 Jupiter Magnetostrictive Transmitters151.6.4.3 Measurement Type: Interface & LevelDisplay Action Comment1StatusIfcLvl% OutputLoopTransmitte

Strona 8 - Wiring Diagram

1646-648 Jupiter Magnetostrictive Transmitters1.6.4.3 Measurement Type: Interface & Level (cont.)Display Action Comment24New PassXxxEnter new pass

Strona 9 - 1.6.1 Operating Parameters

46-648 Jupiter Magnetostrictive Transmitters171.6.4.4 Measurement Type: Level & InterfaceDisplay Action Comment1StatusLevel% OutputLoopTransmitter

Strona 10

1846-648 Jupiter Magnetostrictive Transmitters1.6.4.4 Measurement Type: Level & Interface (cont.)Display Action Comment24New PassXxxEnter new pass

Strona 11

46-648 Jupiter Magnetostrictive Transmitters191.7 Configuration Using HARTA HART (Highway Addressable Remote Transducer)remote unit, such as a HART 37

Strona 12

46-648 Jupiter Magnetostrictive TransmittersRead this Manual Before InstallingThis manual provides information on the Jupiter®magne-tostrictive transm

Strona 13

2046-648 Jupiter Magnetostrictive Transmitters1.7.4 HART Menu (Jupiter 2.0)1 Calibration2 Basic Setup3 Advanced Setup4 Diagnostics5 Review1 Device Set

Strona 14

46-648 Jupiter Magnetostrictive Transmitters211.8 FOUNDATION Fieldbus Digital Communications1.8.1 DescriptionFOUNDATION Fieldbus™is a digital communic

Strona 15

2246-648 Jupiter Magnetostrictive TransmittersDetails regarding cable specifications, grounding, termination,and other network information can be foun

Strona 16

46-648 Jupiter Magnetostrictive Transmitters23An important requirement of Fieldbus devices is the inter-operability concept mentioned above. Device De

Strona 17

2446-648 Jupiter Magnetostrictive Transmitters2.0 Reference InformationThis section presents an overview of the operation of theJupiter magnetostricti

Strona 18

46-648 Jupiter Magnetostrictive Transmitters252.3.1 TroubleshootingProblem SolutionTransmitter does not track level (External Mount) Remove transmitte

Strona 19 - 1.7.3 HART Revision Table

2646-648 Jupiter Magnetostrictive Transmitters2.3.2 Status MessagesDisplay Message Action CommentOK None Normal operating modeInitial None Shown at po

Strona 20 - Measurement Type

46-648 Jupiter Magnetostrictive Transmitters27Unit Equippedwith Display?Check Voltage atTerminal BoardNo Voltage?Check PowerSupply and WiringNegativeV

Strona 21 - 1.8.1 Description

2846-648 Jupiter Magnetostrictive TransmittersCheck LoopCurrent at TerminalBoardLoop CurrentAbove 23 mA?Remove ElectronicModuleLoop CurrentAbove 1 mA?

Strona 22 - 1.8.3 Device Descriptions

46-648 Jupiter Magnetostrictive Transmitters29Unit with DisplayCommunicatesErrorDisplayed?No Level SignalError CodesNo NoYesReload allParametersYesYes

Strona 23 - 1.8.4 Intrinsically Safe

46-648 Jupiter Magnetostrictive TransmittersTable of Contents1.0 Installation ...41.

Strona 24 - D EINUSA

3046-648 Jupiter Magnetostrictive Transmitters2.4 Agency Drawing/Specifications2.4.1 Agency Drawing

Strona 25 - 2.3 Troubleshooting

46-648 Jupiter Magnetostrictive Transmitters312.4.1 Agency Drawing

Strona 26 - 2.3.2 Status Messages

3246-648 Jupiter Magnetostrictive Transmitters2.5 Maintenance2.5.1 Keep Control CleanPeriodic inspections are a necessary means to keep yourlevel cont

Strona 27

46-648 Jupiter Magnetostrictive Transmitters33➀➁➂➂➃➄Replacement Parts Diagram2.6 Replacement Parts2.6.1 Parts IdentificationItem Description Part Numb

Strona 28

3446-648 Jupiter Magnetostrictive Transmitters2.7 Specifications2.7.1 PerformanceAccuracy ±0.015"Repeatability ±0.005% of full span or 0.005"

Strona 29

46-648 Jupiter Magnetostrictive Transmitters352.7.2 FunctionalMeasured variables Continuous liquid levelInput power (at terminals) 12-28 VDCSignal out

Strona 30 - 2.4.1 Agency Drawing

3646-648 Jupiter Magnetostrictive TransmittersDirect Insertion Top Mount Top Mount Offset Top Mount OffsetHigh Temperature BendBottom Mount Offset Bot

Strona 31

46-648 Jupiter Magnetostrictive Transmitters37Accuracy The maximum positive and negative % devia-tion from the actual value over the total span.ANSI A

Strona 32 - 2.5.1 Keep Control Clean

3846-648 Jupiter Magnetostrictive TransmittersNonhazardous Area An area where no volatile mixturesof vapors/gas and oxygen will be found at any time.

Strona 33 - OUNDATION fieldbus

46-648 Jupiter Magnetostrictive Transmitters39Copy blank page and store calibration data for future reference and troubleshooting.Item Value Value Val

Strona 34 - 2.7 Specifications

446-648 Jupiter Magnetostrictive Transmitters1.0 InstallationCaution: If equipment is used in a manner not specified by manu-facturer, protection prov

Strona 35 - OUNDATION Fieldbus output

Service PolicyOwners of Magnetrol/Orion Instruments controls mayrequest the return of a or any part of an instrument forcomplete rebuilding or replace

Strona 36 - Inches (mm)

46-648 Jupiter Magnetostrictive Transmitters51.3 Before You BeginCaution: This instrument is intended for use in InstallationCategory II, Pollution De

Strona 37 - Glossary

646-648 Jupiter Magnetostrictive Transmitters1.3.3 Operational ConsiderationsExterior ambient temperature of the service should notexceed the design s

Strona 38

46-648 Jupiter Magnetostrictive Transmitters73. Attach the lower clamp and tighten so that it remains inplace, but loose enough so that there is still

Strona 39 - Transmitter

846-648 Jupiter Magnetostrictive Transmitters1.5 WiringCaution: The HART version Jupiter transmitter operates at 12–28VDC. The Fieldbus version operat

Strona 40 - SUPERSEDES: November 2007

46-648 Jupiter Magnetostrictive Transmitters91.6 Configuring the TransmitterThe Jupiter transmitter comes configured from the factorywith regard to pr

Komentarze do niniejszej Instrukcji

Brak uwag